DNA/RNA concentration and purity (A260)
Absorbance at 260 nm to a concentration (ng/µL), and purity read from the 260/280 and 260/230 ratios, with their thresholds.
Spectrum
three measurements- Calculation
- C = A260 × 50 × 1 = 1 × 50 × 1 = 50 ng/µL
- Ratios
- A260/A280 = 1 / 0.556 = 1.7986
- A260/A230 = 1 / 0.476 = 2.1008
Notes (2)
The calculation uses only the three wavelengths entered, with no interpolation.
The grey band is the reference range for the classification; outside it does not mean unusable.
Result
continuous calculation- 260/280 ratioA260/A280
- 1.7986 (expected)
- 260/230 ratioA260/A230
- 2.1008 (expected)
- Conversion factorf
- 50 µg/mL
- Dilution factord
- 1
The undiluted sample contains 50 ng/µL.
Calculation, assumptions and sources
OverviewOne absorbance unit at 260 nm corresponds to 50 µg/mL of double-stranded DNA, 40 for RNA and 33 for single-stranded DNA; the concentration accounts for the dilution factor.
One absorbance unit at 260 nm corresponds to 50 µg/mL of double-stranded DNA, 40 for RNA and 33 for single-stranded DNA; the concentration accounts for the dilution factor. The 260/280 and 260/230 ratios flag contaminants, with the published thresholds (1.8 for DNA, 2.0 for RNA). A ratio without a measured absorbance at 280 or 230 nm is not assessed.
Method & formulasC (µg/mL) = A260 × factor × dilution
Beer-Lambert law at 260 nm, in a 1 cm cuvette: conventional factors convert absorbance into mass concentration. ng/µL and µg/mL are the same unit.
- A260
- the absorbance at 260 nm, the absorption maximum of nucleic bases.
- 260/280 ratio
- sensitive to proteins and phenol, which absorb at 280 nm. Expected around 1.8 for DNA, 2.0 for RNA.
- 260/230 ratio
- sensitive to chaotropic salts, EDTA and phenol, which absorb at 230 nm. Expected between 2.0 and 2.2.
- Dilution factor
- the ratio between the sample and the solution measured. Diluted tenfold, the true concentration is ten times the reading.
Validity domainThe factors assume an average base composition and a 1 cm cuvette; for a short oligonucleotide, the extinction coefficient of the sequence is more accurate.
The factors assume an average base composition and a 1 cm cuvette; for a short oligonucleotide, the extinction coefficient of the sequence is more accurate. Photometry is reliable between about 0.1 and 1 absorbance unit. Above, dilute; below, noise dominates and a fluorometric intercalating-dye assay is better suited.
Purity ratios and contaminantsA low 260/280 flags residual proteins from a phenol-chloroform extraction; a low 260/230, the guanidine salts of column kits or the EDTA of the buffer.
A low 260/280 flags residual proteins from a phenol-chloroform extraction; a low 260/230, the guanidine salts of column kits or the EDTA of the buffer. Remedies: rewash the column, reprecipitate, or change the blank. Good ratios do not guarantee integrity; free nucleotides and degraded DNA also absorb at 260 nm, and only electrophoresis checks it.